Geko™ FAQs
What is the geko™ device?
What does it do?
How does it work?
Who uses it?
Where is it worn?
The geko™ device is intended for use only on the lower limbs. Secure the device to the leg so that the location marker (the line of arrows) is positioned directly over the centre of the fibula head. When correctly fitted, the longer tail should wrap around to the back of the leg, below and parallel to the crease of the knee.
This positioning ensures the negative electrode is placed over the common peroneal nerve, with the ‘plus’ button at the front of the leg and the ‘minus’ button at the rear. The device can also be applied in two alternative locations: just above or just below the crease of the knee, aligned with the lateral outer tendon.
How long should it be worn?
The healthcare professional prescribing the geko™ device will provide guidance on the wear protocol based on your condition. For example, to prevent venous thromboembolism, the device should be worn on both legs for 24 hours per day during the acute phase (the first 1 to 6 days).
Why is there a limited run-time on the T device?
What’s the correct stimulation level?
Is it painful?
Can I sleep with it on?
Is it waterproof?
Single-use or reusable?
Hospital Application
The geko™ is a single-use medical device, meaning it is designed for use on a single patient as part of a single course of treatment.
Wound Therapy
For wound therapy, the geko™ device should be used for a maximum of two days before being discarded. The device can be stored on the backing card between uses. Please refer to the Instructions for Use (IFU) for further details.
How is it different from other electrostimulation devices?
NMES vs TENS—what’s the difference?
What pulse does it use?
The geko™ device delivers a constant current, square wave, asymmetrical, charge-balanced pulse. The term “constant current” refers to the device’s ability to maintain a consistent output during the duration of the pulse. The device’s voltage adjusts according to the patient’s electrical resistance to ensure the required current output is achieved.
The stimulation pulse is described as a square wave, meaning the current rapidly rises to the desired value, remains constant for the duration of the pulse, and then quickly drops back to zero. This creates a rectangular or square shape when viewed on an oscilloscope, hence the name “square wave.”
Following the primary stimulation pulse, there is a return pulse of lower intensity but longer duration, with opposite polarity (asymmetrical) to ensure the total electrical charge is zero (charge-balanced). This charge-balancing feature helps prevent electrochemical changes in the skin.
How much blood-flow increase is seen?
What about pacemakers or other implants?